Optical Unit
DX6106 Series
User Guide
Joint Stock Company
RMT Ltd
RMT Ltd
2002
Rev. 1.10
Vista de pagina 0
1 2 3 4 5 6 ... 60 61

Indice de contenidos

Pagina 1 - User Guide

Optical UnitDX6106 SeriesUser GuideJoint Stock CompanyRMT LtdRMT Ltd2002Rev. 1.10

Pagina 2

Design FeaturesThe DX6106 Optical Unit is specially designed for a fastresponse, high sensitivity, low noise and low powerconsumption.A number of desi

Pagina 3

2-5Theory of OperationRMT LtdRMT LtdRMT LtdRMT LtdDX6106The subsequent description assumes User has alreadydeveloped, manufactured and connected to th

Pagina 4

44332210YAYAYAYAAX ×+×+×+×+=DDY0=REV. 1.10/2002RMT LtdRMT LtdRMT LtdRMT LtdDX61062-6Calculated coefficients … of polynomialexpression and zero ratio m

Pagina 5

Theory of OperationRMT LtdRMT LtdRMT LtdRMT LtdDX61062-73. To preserve high accuracy of the device it isnecessary to do “zero” adjustments periodicall

Pagina 6

REV. 1.10/2002RMT LtdRMT LtdRMT LtdRMT LtdDX61062-8

Pagina 7

Construction of Optical UnitRMT LtdRMT LtdRMT LtdRMT LtdDX61063. Construction of Optical Unit3-1The DX6106 OpticalUnit (Fig. 3.1 and 3.2)consists of a

Pagina 8

REV. 1.10/2002RMT LtdRMT LtdRMT LtdRMT LtdDX61063-2The body of gassampling cell is made ofanodized aluminum alloy(Fig. 3.3). It has two gasinlets with

Pagina 9

Construction of Optical UnitRMT LtdRMT LtdRMT LtdRMT LtdDX61063-3The main parameters of gas sampling cell are in Table3.1Table 3.1. Main parameters of

Pagina 10 - RMT LtdRMT Ltd

REV. 1.10/2002RMT LtdRMT LtdRMT LtdRMT LtdDX61063-4Fig. 3.6 DX6106.C40 gas sampling cell outline dimensions(in millimeters)2742402736.5The outline dim

Pagina 11

Construction of Optical UnitRMT LtdRMT LtdRMT LtdRMT LtdDX61063-5OptocomponentPhotodetectorMeasuringchannelemitterReferencechannelemitterFig. 3.7. OPR

Pagina 12

CopyrightLimited WarrantyTrademark AcknowledgmentsAll right reserved. Reproduction in any manner, in whole or in partis straightly prohibited without

Pagina 13

REV. 1.10/2002RMT LtdRMT LtdRMT LtdRMT LtdDX6106As an example, the parametersof optoelements for methane(CH ) measurements areintroduced in the table

Pagina 14

Construction of Optical UnitRMT LtdRMT LtdRMT LtdRMT LtdDX6106Fig. 3.10. 6102OptocomponentMating Module6102 Optocomponent Mating ModuleThe 6102 Optoco

Pagina 15

Fig. 3.13. Functional Diagram of 6102 OptocomponentREV. 1.10/2002RMT LtdRMT LtdRMT LtdRMT LtdDX6106Legend– MOSFET switch– amplifier– Light Emitter– Ph

Pagina 16

Construction of Optical UnitRMT LtdRMT LtdRMT LtdRMT LtdDX6106Table 3.2. DX6106 s pinsfunction descriptionystem interface connector X1Pin Mnemonic Des

Pagina 17

REV. 1.10/2002RMT LtdRMT LtdRMT LtdRMT LtdDX6106PreamplifierThe preamplifier alternatively processes the signals ofmeasuring and reference channels.Th

Pagina 18

Construction of Optical UnitRMT LtdRMT LtdRMT LtdRMT LtdDX6106Within the gas concentration varying scale, the outputsignal of measuring channel change

Pagina 19

REV. 1.10/2002RMT LtdRMT LtdRMT LtdRMT LtdDX6106MTB=20HSCH1 = .5VFig. 3.16. Typical output signalwaveformThe preamplifier typical output signal wavefo

Pagina 20

Construction of Optical UnitRMT LtdRMT LtdRMT LtdRMT LtdDX6106Thermoelectric CoolerThe TEC circuit diagram ispresented in Fig. 3.17.Driving by TE cool

Pagina 21

REV. 1.10/2002RMT LtdRMT LtdRMT LtdRMT LtdDX6106Fig. 3.18. TEC power consumption vs operatingtemperaturePower, W0.51.01.52.00–15–20 –10–50510Temperatu

Pagina 22

Construction of Optical UnitRMT LtdRMT LtdRMT LtdRMT LtdDX6106consumption; at higher temperature the output signals(and signal/noise ratio) are lower.

Pagina 23 - Pin DescriptionMnemonic

ContentsRMT LtdRMT LtdRMT LtdRMT LtdDX6106C-11. Introduction 1-12. Theory of Operation3.Principles of Operation 2-1Design Features 2-4Operation Overvi

Pagina 24

REV. 1.10/2002RMT LtdRMT LtdRMT LtdRMT LtdDX6106Fig. 3.21. Maximal temperaturedifference vs operating currentfor 1MT04-059-16.(At zero heat load)If st

Pagina 25

Construction of Optical UnitRMT LtdRMT LtdRMT LtdRMT LtdDX6106T = +20°CAMBTE Cooler SpecificationsElectrical ParametersDynamical Parameters26.80.512.3

Pagina 26

REV. 1.10/2002RMT LtdRMT LtdRMT LtdRMT LtdDX61063-18

Pagina 27

Construction of Optical UnitRMT LtdRMT LtdRMT LtdRMT LtdDX6106ThermistorFig. 3.22. Thermistorconnection in the 6102RTUREFRLCt°X17X13X16X14X18OUTTGNDAG

Pagina 28

REV. 1.10/2002RMT LtdRMT LtdRMT LtdRMT LtdDX6106Fig. 3.23. Calibration of thermistor vs measured temperature253020151050–40 –20 0 20 40Temperature, °C

Pagina 29

Construction of Optical UnitRMT LtdRMT LtdRMT LtdRMT LtdDX6106One can see that the temperature measurement accu-racy depends directly on . So the volt

Pagina 30

REV. 1.10/2002RMT LtdRMT LtdRMT LtdRMT LtdDX6106It is obvious that the scheme responds to temperaturein a nonlinear manner. At over a limitedtemperatu

Pagina 31 - Min Typ Max

Construction of Optical UnitRMT LtdRMT LtdRMT LtdRMT LtdDX6106Fig. 3.25. Typical thermistor usageDX6106 Optical UnitADCRTRLCt°X17X13X16X14X18FilterX12

Pagina 32

REV. 1.10/2002RMT LtdRMT LtdRMT LtdRMT LtdDX6106Light EmittersElectronicscheme for LightEmitters drivingis shown in Fig.3.26.The MOSFETtransistors are

Pagina 33

Construction of Optical UnitRMT LtdRMT LtdRMT LtdRMT LtdDX6106Power Supply 5 V(+4…+6 V areavailable) throughresistor charges thecapacitor in timedurat

Pagina 34

C-2REV. 1.10/2002RMT LtdRMT LtdRMT LtdRMT LtdDX6106

Pagina 35

REV. 1.10/2002RMT LtdRMT LtdRMT LtdRMT LtdDX6106Fig. 3.28. Typical LED control circuitDACDACMicrocontroller+X26X24X25X23X21RECRSRGRGDX6106 Optical Uni

Pagina 36

Construction of Optical UnitRMT LtdRMT LtdRMT LtdRMT LtdDX6106is easy to plot the load line (the curve in Fig.3.X)and to see that the forward current

Pagina 37

REV. 1.10/2002RMT LtdRMT LtdRMT LtdRMT LtdDX6106The standard ElectricallyErasable PROM (EEPROM)24LC64 chip with two wire serialis placed on OpticalUni

Pagina 38

Construction of Optical UnitRMT LtdRMT LtdRMT LtdRMT LtdDX6106The detailed description of 24LC64 transfer protocol ispossible to obtain, for example,

Pagina 39 - T = +25°C

3-30REV. 1.10/2002RMT LtdRMT LtdRMT LtdRMT LtdDX6106EEPROM Data FormatVarious operating parameters can be stored in on-board EEPROM circuit:See recomm

Pagina 40 - Microcontroller

3-31Construction of Optical UnitRMT LtdRMT LtdRMT LtdRMT LtdDX6106*)used only with DX6101 ControllerTable 3.4. EEPROM Data Format891076Parameters of t

Pagina 41

3-32REV. 1.10/2002RMT LtdRMT LtdRMT LtdRMT LtdDX6106

Pagina 42

4. Installation TipsThere are two external power sources required forOptical Unit supply.An external +5V and -5V DC power sources with±2.5% output tol

Pagina 43 - Parameter Value

4-2REV. 1.10/2002RMT LtdRMT LtdRMT LtdRMT LtdDX6106Fig. 4.1. DX6106 power supplies connectionX22X26X15X17X13X16X14X18X12X11X24X25X23X21Powersupply+5VL

Pagina 44

5-1CalibrationRMT LtdRMT LtdRMT LtdRMT LtdDX61065. CalibrationPreparationFirst of all User should prepare the set of calibrationgases.The number of ca

Pagina 45

1-1IntroductionRMT LtdRMT LtdRMT LtdRMT LtdDX61061. IntroductionThe company RMT Ltd introduces new DX6106 seriesof Optical Units for gas measurement s

Pagina 46

Any other sets of standard samples User can applyaccording to own reasons. But be sure that the samplesare within specified measurement range and thec

Pagina 47

CalibrationRMT LtdRMT LtdRMT LtdRMT LtdDX61065-3Zero AdjustmentsTo ensure the high accuracy, simple adjustment can bemade during operation to adjust “

Pagina 48 - Shielding Ground

Re-CalibrationIn standard option, the DX6106 Optical Unit is deliveredwith one calibration data. The calibration is made atoptimal operating temperatu

Pagina 49 - Standard Kit

MaintenanceRMT LtdRMT LtdRMT LtdRMT LtdDX61066-16. MaintenanceOptics CleaningIf the output signal of either channels (or both) becameappreciably less

Pagina 50

6-2REV. 1.10/2002RMT LtdRMT LtdRMT LtdRMT LtdDX6106Clean nowthe window of the optopair,spherical mirror,internal mirror (with a hole) from both sides.

Pagina 51

Standard KitRMT LtdRMT LtdRMT LtdRMT LtdDX61067-17. Standard Kit146325798Optical UnitConnector’s housingConnector’s shrouded headerInterconnection cab

Pagina 52

RMT LtdRMT LtdRMT LtdRMT LtdDX61067-2REV. 1.10/2002

Pagina 53

SpecificationsRMT LtdRMT LtdRMT LtdRMT LtdDX61068-1CO2mnCHMeasured gasesCarbon DioxideHydrocarbons8. SpecificationsTypeDetectorNDIR multipass schemeLe

Pagina 54

REV. 1.10/2002RMT LtdRMT LtdRMT LtdRMT LtdDX61068-20...1000 ppm< 3 ppm10 ppm0...5 % vol< 0.15 %0,5 %0.02 %0...20 % vol< 0.15 %0,5 %0...1000 p

Pagina 55

9. Ordering GuideExample:[DX6106][01][504][A]DX6106 Series,CO Gas Option,0...5.0 10 ppm (0...5%) concentration range,aspiration type gas cell,2×4DX610

Pagina 56

REV. 1.10/2002RMT LtdRMT LtdRMT LtdRMT LtdDX61061-2

Pagina 57

RMT LtdRMT LtdRMT LtdRMT LtdDX61069-2REV. 1.10/2002

Pagina 58

RMT LtdRMT LtdRMT LtdRMT LtdDX6106SpecificationsRMT Ltd. Leninskij prosp. 53, Moscow 119991 Russiaphone: 095-132-6817 fax: 095-135-0565e-mail: rmtcom@

Pagina 59 - XX XXXXX . XX . XXX

2-12. Theory of OperationPrinciples of OperationThe NDIR (Non-Dispersive Infra-Red Spectroscopy)measurement method is implemented in the DX6106Optical

Pagina 60

2-2where– intensity of light before pass through the gasvolume;– intensities of light after pass through the gasvolume;– absorption coefficient of the

Pagina 61

2-3Reference channel isused for indirectmeasuring of theinitial intensity oflight, and allows toeliminate actualmeasurementsconditions (totaltranspare